选择您所在的国家/地区
KFD0-TR-Ex1 RTD Converter
主要优势一览
- 1-channel isolated barrier
- 24 V DC supply (loop powered)
- 2- or 3-wire Pt100 RTD input
- Output 4 mA ... 20 mA, temperature linearization selectable
- DIP switch selectable ranges
- Sensor breakage detection
显示详情






注意
在转换前以倍加福有限公司或倍加福股份公司名义发布的所有产品文档(如证书、符合性声明等)也适用于倍加福欧洲公司。
EN
- EN
- DE
- ES
- FR
- PL
KFD0-TR-Ex1 技术数据
| Signal type | Analog input |
| Rated voltage | Ur | 12 ... 35 V DC loop powered |
| Power dissipation | 0.4 W |
| Connection side | field side | |
| Connection | terminals 1, 2-, 3+ suitable for Pt100, 2- and 3-wire connection | |
| Lead resistance | max. 100 Ω per line | |
| Measuring current | approx. 1 mA |
| Connection side | control side | |
| Connection | terminals 9+, 8- | |
| Load | (U -12 V) / 0.02 A | |
| Current output | 4 ... 20 mA , limited to ≤ 35 mA | |
| Fault signal | upscaling ≥ 22 mA (limited to 35 mA) |
| Measuring range | fn | span without linearization 25 ... 800 °C (77 ... 1472 °F)/with linearization 25 ... 375 °C (77 ... 707 °F), both adjustable zero point without linearization -200 ... 400 °C (-328 ... 752 °F)/with linearization -30 ... 375 °C (-22 ... 707 °F), both adjustable | |||||||||
| Deviation | |||||||||||
| |||||||||||
| Rise time | 250 ms | ||||||||||
| Input/Output | available |
| Control elements | DIP switch rotary switch | |
| Configuration | via DIP switches via rotary switch | |
| Labeling | space for labeling at the front |
| Electromagnetic compatibility | |||||
| |||||
| Insulation coordination | EN 50178 | |
| Galvanic isolation | EN 50178 | |
| Degree of protection | IEC 60529 |
| Ambient temperature | -20 ... 60 °C (-4 ... 140 °F) |
| Degree of protection | IP20 | ||||||||||
| Connection | screw terminals | ||||||||||
| Mass | approx. 150 g | ||||||||||
| Dimensions | 20 x 119 x 115 mm (0.8 x 4.7 x 4.5 inch) (W x H x D) , housing type B2 | ||||||||||
| |||||||||||
| Mounting | on 35 mm DIN mounting rail acc. to EN 60715:2001 | ||||||||||
| EU-type examination certificate | ZELM 00 ATEX 0036 | |||||||||||||
| ||||||||||||||
| Certificate | TÜV 01 ATEX 1777 X | |||||||||||||
| ||||||||||||||
| Galvanic isolation | ||||||||||||||
| ||||||||||||||
| Directive conformity | ||||||||||||||
| ||||||||||||||
| CSA approval | 1029981 | |||||||
| ||||||||
| IECEx approval | ||||||||
| ||||||||
| Supplementary information | Observe the certificates, declarations of conformity, instruction manuals, and manuals where applicable. For information see www.pepperl-fuchs.com. |